Group code f-e-e17e0d8d6c-:
“power supply (PSU) communication fail”
Situation
Solution
Only one PSU has a communication fail.
Check if the PSU is correctly inserted in the PSU power rack.
Check power cables between PSU and power rail in the mains input
compartment (reference 1 image and
reference 2 image)
Check if the PSU is powered. Ensure that the three phases of
the mains power is correctly wired.
Multiple PSU have a communication fail.
Check if the communication wire between the PSU power rack (RJ45
plug, reference 5 image) and Laser
Control board (10 pole connector, reference 6 image) is properly inserted.
Check if the communication wire between the PSU power racks
is properly inserted (reference 3 and 4 image). (only applicable in case two power
racks are installed).