Group code f-e-e17d14e4d6c-:
“power supply write on fail”
Situation
Solution
Only one PSU has a communication fail.
Check if the PSU is correctly inserted in the PSU power rack.
Check power cables between PSU and power rail in the mains input
compartment (reference 1 image and reference
2 image)
Check if the PSU is powered. Ensure that the three phases of
the mains power is correctly wired.
Multiple PSU have a communication fail.
Check if the communication wire between the PSU power rack (RJ45
plug, reference 5 image) and Laser Control
board (10 pole connector, reference 6 image)
is properly inserted.
Check if the communication wire between the PSU power racks
is properly inserted (reference 3 and 4 image).
(only applicable in case two power racks are installed).