Situation |
Solution |
|---|---|
|
The liquid cooling circuit of the Light Processor is mistakenly excluded from the main liquid cooling circuit. Most likely the other DMD temperatures are too high as well. |
|
|
Malfunction small Airdryer. The other DMD temperatures are too high as well (only for 45L or 60L projector model) |
Check for warnings and errors related to the small Airdryer and resolve these first. |
|
Malfunction Heat Exchanger. The other DMD temperatures are too high as well (only for 22L or 30L projector model) |
|
|
Malfunction fan of the DMD. |
Check for fan errors related to this DMD and resolve these first. |
|
Defect temperature sensor. |
Replace the whole Light Processor Unit. Contact Barco for further instructions to repair the malfunction Light Processor Unit. It is too critical to replace this temperature sensor in the field. The replacement of this temperature sensor must be done at factory or at an authorized service center. See service procedure "Light Processor replacement process". |