Situation |
Solution |
|---|---|
|
Malfunction Fan Control board or SMPS board. |
Measure on the Signal Backplane the +24V voltage on pin 9, 10, 11 or 12 of the connector of the black wire unit which comes from the SMPS board. See image. If the measured voltage is about 24V then replace the Fan Control board. See "Replacement of a Card Cage board". Otherwise replace the SMPS board. See "Replacement of the SMPS board". |
|
Damaged insulation of the wire unit (reference 1 image) of the Cathode fan (reference 2 image). |
Check the insulation of the wire unit of the Cathode fan. The Cathode fan is located at the right inner side of the Lamp compartment.
|
|
Malfunction Signal Backplane (bad connection) |
Replace the Signal Backplane. |
![]() |
![]() |