Code 20131: “LCB – Turn-on LCM 1 failed”

Situation

Solution

Light Source cooling could not be switched on during startup sequence

  • Check / reseat the communication cable between the LCB (J600) and the light source fan interconnection board (J101).

  • Upgrade the software to the latest version. See procedure "Software update via Communicator (DC update companion)".

  • If problem persists, replace the fan interconnection board. See service procedure "Replacing the Light Source Fan Interconnection Board".