With the Communicator, go to → → →
The following errors can occur:
|
Error |
Solution |
|---|---|
|
Link decryptor - User loader integrity error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - Main application integrity error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - RNG hardware integrity error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - DRNG hardware integrity error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - RSA algorithm integrity error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - AES algorithm integrity error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - HMAC algorithm integrity error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - SHA algorithm integrity error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - TLS integrity error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - FPGA configuration integrity error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - FPGA cinelink 2 decryption integrity error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - Real time clock error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - FPGA configuration error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - FPGA temperature out of range |
Temperature on the link decryptor is to high. Check for other fan or temperature error messages. Check if the front cover of the HDSDI module is closed. |
|
Link decryptor - RNG Hardware duplicate output error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - DRNG Hardware duplicate output error |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - 1.20V supply out of range |
Replace SMPS. See service manual chapter "Switch Mode Power Supply (SMPS)". Replace Signal backplane. See service manual chapter "Signal Backplane replacement process". Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - 1.80V supply out of range |
Replace SMPS. See service manual chapter "Switch Mode Power Supply (SMPS)". Replace Signal backplane. See service manual chapter "Signal Backplane replacement process". Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - 2.50V supply out of range |
Replace SMPS. See service manual chapter "Switch Mode Power Supply (SMPS)". Replace Signal backplane. See service manual chapter "Signal Backplane replacement process". Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - 3.30V regulator out of range |
Replace SMPS. See service manual chapter "Switch Mode Power Supply (SMPS)". Replace Signal backplane. See service manual chapter "Signal Backplane replacement process". Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - Security tamper |
The link decryptor has been tampered. Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - Top side security enclosure open |
The top side enclosure has been tampered. Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - Bottom side security enclosure open |
The bottom side enclosure has been tampered. Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - Software command Zeroization |
Replace link decryptor. See service manual chapter "Replacement of the Link Decryptor". |
|
Link decryptor - Physical marriage tamper |
The ICP and/or link decryptor have been removed and inserted again. You need to remarry both. Start the clear security warning procedure. See service manual chapter "Authorization to clear security warning on the projector". |
|
Link decryptor - Logical marriage tamper |
The ICP and link decryptor are not a couple. This means one of both has been replaced. You need to marry both. Start the clear security warning procedure. See service manual chapter "Authorization to clear security warning on the projector". |
|
Link decryptor - Service door tamper |
The Card Cage cover has been removed. Clear the security warning. See service manual chapter "Authorization to clear security warning on the projector". The cover plate of the Light Processor compartment has been removed. Clear the security warning. See service manual chapter "Authorization to clear security warning on the projector". Defect security switches. If all compartments and devices are installed and security error will not authorize, replace DCI security switches. Malfunction Signal Backplane. If switch replacement is not the solution, then replace Signal Backplane. |
|
Link decryptor - Service log error |
The Security Log Error indicates that there is no more room to write log entries in the link decryptor log file. When this error is active, the DCI compliant Server needs to extract the log data (via ASM/TLS session) from the link decryptor Security Log. Reads of the security log by any other entity/means will not impact this error. This error will remain active until the Server reads enough log entries to create room for 512 log entries. This can take some time. Playback will be prohibited (black image) as long as this error is active. |
|
Link decryptor - Security battery low warning |
The battery of the link-decryptor is low. Leave the projector on for some time to recharge the battery. |
|
Link decryptor - Security log warning |
The security log warning indicates that there are less than 512 log entry locations available in the link decryptor log before the log is full. The attached server needs to read the log which will clear the log of the projector. When this warning is active, the DCI compliant server needs to extract the log data (via ASM/TLS session) from the link decryptor Security Log. This can take some time. Reads of the security log by any other entity/means will not impact this warning. |