Situation
Solution
The wire unit (reference C ) of the fan (reference FL ) is disconnected.
Reconnect the wire unit of the cathode fan. The cathode fanis located between the lamp compartment and the Input & CommunicationUnit.
Wire unit of the fans and motor units (reference W ) disconnected from the Power Backplane.
Remove the Input & Communication unit and check if the wireunit is inserted in the Power Backplane.
Blocked fan.
Unblock the fan. Ensure the fan can turn freely.
Damaged wire unit of the fan.
Check if the wire unit of the fan is not damaged. Repair ifpossible, otherwise replace with new one.
Fan end of life.
Replace the fan.
Malfunction Fan Control Board
Replace the Fan Control Board.
Other fans are also not spinning.
Replace the SMPS Board.
Defect Power Backplane
Replace the Power Backplane.
Fan LampInfo Module at the rear of the lamp house.
Connectionfan wire unit.