Code 5053: “enginefan - speed low” (Warning)

Situation

Solution

Wire unit (reference C ) of the fan (reference F3) in the compartment of the Light ProcessorUnit is disconnected. Note that this fan is located between the compartmentof the Input & Communication Unit and the Compartment of the LightProcessor Unit.

Check the connection of the wire unit of the fan. The fan isaccessible by removing a small cover plate inside the compartmentof the Input & Communication Unit.

Wire unit of the fans and motor units (reference W ) disconnected from the Power Backplane.

Remove the Input & Communication unit and check if the wireunit is inserted in the Power Backplane.

Blocked fan.

Unblock the fan. Ensure that the fan can turn freely.

Damaged wire unit.

Check if the wire unit of the fan is not damaged. Repair ifpossible, otherwise replace with new one.

Fan end of life.

Replace the fan. See service manual chapter Light Processor fan.

Malfunction Fan Control Board

Replace the Fan Control Board.

Other fans are also not spinning.

Replace the SMPS Board.

Defect Power Backplane

Replace the Power Backplane.